The 13C calibration is done with a sample of acetate 13C labeled on carbon 2 (methyl group). The acetate is dissolved in deuterated water.
Both reference samples may be found in Wilmad tubes (round bottom NMR tubes) found in the NMR room.
To reroute the 15N RF through the preamplifier and to the probe in the avance 600 also requires a hardware change. This involves switching the cable (and accompanying filter) which connects the front of the X-nucleus preamp (the second one down in the stack) to the probe. Normally a 13C bandpass filter is attached to the front of the X-nucleus preamp which in turn is connected to the 13C channel on the probe by a cable. This should be changed so that the 15N bandpass filter is attached to the front of the preamp which in turn is connected to the 15N channel of the probe. Normally the 15N channel is set up so that the input from the amplifier bypasses the preamp, is passed through the 15N bandpass filter, and then into the 15N channel on the probe. Thus, for the purposes of the 15N tuning experiment, the connection between the 15N amplifier and 15N bandpass filter can be broken and in turn the 15N bandpass filter attached to the front of the X-nucleus preamp. It's OK if the 13C connection is broken during the 15N tuning operation). Two important things to remember are to a) Make sure you change the cable back after you tune the 15N probe [including before running calib15n], and b) be very gentle with the 15N and 13C bandpass filters; they contain inductors (essentially small coils of copper wire) that are sensitive to mechanical distortion which in turn will alter their filtering properties.
The Avance 500 and 700 spectrometers will have two X-nucleus preamplifier units each to avoid having to change cables.
As soon as the antiphase state is reached, the 90 degree 15N or 13C pulse to be calibrated is delivered. If the pulse is correct, it will convert the 2IySz state to the multiple quantum, 2IySy state. This is called a multiple quantum state because there is transverse spin (x or y) on both of the coupled atoms. In accordance with quantum theory, multiple quantum states do not produce a signal. The calibration experiment consists of varying the 15N or 13C pulse time to make the antiphase signal disappear. Pulse times past 90 degrees cause the antiphase signal to reappear with both peaks inverted relative to the 0 pulse time antiphase pattern. A 180 degree pulse causes the inverted antiphase pattern to reach a maximum intensity.
Check the pulse program for which the pulse is intended for information about the pulse. It may be shaped or square, and a nominal power and pulse time will be recommended. Check the comments in the calib15n or calib13c pulse program to see which pulse and power setting are to be used for the 13C or 15N pulse. For those who learned on the AMX 500, the Avance systems have some changes in the names of parameters. Power attenuation levels associated with square pulses p1, p2, etc. are now known as PL1, PL2, ... (not HL1, HL2, DL1, DL2, DBL0,DBL1, etc.). [Note: the actual commands are in lower case.] Attenuation levels are on a scale that can be either positive or negative, rather than starting at 0. AMX power levels vary between 0 and + 90 dB; Avance power levels vary between -6 and +120 dB). Nominal power levels recommended for the AMX 500 will NOT be correct for the Avance 500 (or any of the Avance spectrometers). Parameters related to shaped pulses start with sp instead of tp. Also note from the pulse program of the experiment into which the pulse will be inserted whether the time is to be adjusted or the power level. For some kinds of pulses, the time must be exactly as specified and the power level should be adjusted to yield a null (meaning flat) antiphase spectrum. Power levels can be adjusted to 0.01 attenuation units on the Avance systems. If this is not noted in the pulse program, it is typical to leave the power level at the recommended nominal level and adjust only the pulse time. However, if the time increases too much, it is possilbe that the pulse program will not accomidate it. So for an increase of greater than about 10%, it may be safer to adjust the power to keep the time closer to the nominal time. If the time increases greatly (eg 2x), then this indicates a problem of some kind. Sometimes the discrepency is just due to a typo or obsolete documentation. However, if your values are high compared to others you've recently taken, the problem could be poor tuning, or some degradation in the performance of the equipment. If the problem is not resolved by repeating tuning, etc., it should reported. If adjusting power, remember that lower (including more negative) attenuation values mean more power is applied. With more power, less time will be needed to achieve a comparable rotation (e.g.. 90 degrees).
First do the experiment with 0 pulse time. For a shaped pulse,
0 is not a legal time, so use the estimated 180 instead. Transform
with ef. In each case, the large signal you will observe is
from the solvent. Flanking the solvent peak will be a smaller antiphase
doublet (one peak up and one down) that is 90 degrees out of phase with
the solvent. You may need to enlarge the scale to observe the doublet.
Phase the doublet and save the phase [save and return]. Subsequently,
zoom on the doublet, click <dp1>, agree with the 3 popup messages, run
paropt through the 180 degrees estimated from the nominal 90 degree time
to crudely verify the position of the 90 degree pulse. Then run paropt
with finer intervals around the 90 degree point to get a closer determination.
If you scale up the signal at the null point, you will notice a residual
out of phase signal probably originating from an imprecision in the
delay time creating the antiphase signal. In the the 13C
calibration, there is a 3rd out of phase signal between these two originating
from incomplete 13C labeling. Andy prefers to judge the
null by the residual signals appearing symmetric with equal intensity above
and below the baseline. We don't have a good accounting of the tolerance
with which the pulse times are required for the different experiments,
although about 0.5% - 1% might be a good rule of thumb in the absence of
more specific information. If you want to try to set the times more
closely, I recommend reobtaining the data at the best estimate of the null,
and phasing on the residual signals, then save phase. For 13C,
use the middle peak, which will be antiphase to the other two. For
15N,
make the pattern symmetrical. Then take readings through the
null reading by zg and efp. Unless the signal you are
trying to null is completely gone, there will be a small dip in the baseline
either on the inside edges or the outside edges of this phased residual
signal. I notice a disturbance (splitting) in one of the 13C
peaks, and use the other. Also, trying to calibrate precisely by
this method may require more precise shimming if there are shape perturbances
around the base of the peak.
draft: 6/23/02 - Steve Hardies