Avance600
| operator/sample specifics | 1H | 15N
on 15N urea standard |
13C
on 13C acetate standard |
| 6/19/02 was when it was first noticed that the 1H channel was requiring 4-5 lower dB | |||
| Steve Hardies
Ubiquitin standard 50 mM KPi, pH 5.8 T=299K |
HNCACBSE (7/1/02)
8.34u @ -6dB (P1@PL1) Don't use -6 dB in future 2080u @ 36dB (P10@sp1, with seduce1.jc) 40.1u @ 8dB (Pll@PLll and pcpd1 with waltz16) |
HNCACBSE (6/19/02)
T=299K 42.3u; -5 dB (P2@PL2) 181u @7.9dB (pcpd2 @PL12 with waltz16) |
HNCACBSE (6/19/02)
T=299K 14.3u; -3 dB (P3 @PL3) 299u @10.83 dB (pcpd3 @ sp3 with seduce1c6) 249u @9.35 dB (pcpd3 @ sp3 with seduce1c5 calibrated but not used) |
| Steve Hardies | HSQC (6/19/02)
T=299K 45u @-4.4dB (P7@PL7) 181u @7.9dB(pcpd3@PL3) Calibrated, not tested |
HSQC (6/19/02)
T=299K 23.7u @1.45dB (P5@PL5) Calibrated, not tested |
|
| 7/17/02 Andy remade coretab; this may slightly alter hydrogen pulse times. | |||
| Steve Hardies
L1, p190 50 mM NaCl, 30 mM NaPi pH 5.79, T=300K |
7/24/02
Series of calibrated square pulses to test coretab linearity 7.95u @ -6dB
|
||
| Steve Hardies
L1,p190, 50 mM NaCl, 30 mM NaPi, pH=5.79, T=300K |
HNCACBSE (7/24)
8.88u @-5dB (P1@PL1) 2056u @35dB(P10@sp1, with seduce1.jc) 40.7u @8dB (Pll@PLll and pcpd1 with waltz16) |
||
| Steve Hardies
L1, p190 50 mM NaCl, 30 mM NaPi; T=300K |
HSQC (7/29)
8.85 @ -5 dB(P1@PL1) 1020u @35 dB (P2@PL2) 2040u @35 dB(p3@sp0 with seduce1.jc) This set done fairly crudely. |
||
| Andy Hinck
Ubiquitin standard in 50 mM KPi, pH 5.8 T=300K |
CBCACONHSE (7/30)
Wisconsin version 8.5 @ -5.1 dB (P1@PL1) 50u @10 dB (pcpd1@PL11, cpdprg1=dipsi2) |
CBCACONHSE (7/30)
T=300, Wisconsin version 45u @-4.4dB(P2 @PL2) 180u@7.9 dB (pcpd2 @PL12, cpdprg2= waltz16) |
CBCACONHSE (7/30)
T=300, Wisconsin version 48.2@7.7 dB (p3@PL3 ) 21.6u@0.8 dB (P6@PL6) 53.5u @ 8.6 dB (P4@PL4) 250u@9.4 dB (P5@sp5 seduce1.jc spoffs5 20068 ) 53.5u@8.5dB (p7@sp7 spnam7=rect32 spoffs7=18107) 47.8u @1.4dB (p8@sp8 spnam8=rect32 spoffs8=18107) 125u@9.4 dB (pcpd3@sp3 spname3=seduce1.jc spoffs3 = 0; cpdprg3=waltz16sp3) |
| Steve Hardies, L1 p190
50 mM NaCl, 30 mM NaPi pH 5.79, T=300 |
CBCACONHSE (7/31)
Wisconsin version 8.85u @ -5 dB (P1@PL1 ) 47.93u @9.5 dB (pcpd1@PL11) |
||
| Andy Hinck
Ubiquitin standard in 50 mM KPi, pH 5.8, T=300K |
CBCACONH (7/30)
NIH version 8.5u@-5.1 dB (P1@PL1) 1055u @ 36 dB (P2@PL2) 50u@10 dB (P30@PL30) |
CBCACONH (7/30)
NIH version , T=300 40@-5 dB (P17@PL17) 170@7.9dB (P31@PL31) |
CBCACONH (7/30)
NIH version, T=300 47.9u@7.6dB (P4@PL4) 53u@8.5 dB (P5@PL5) 21.4u@0.7 dB(P6@PL6) 47.4u @1.4 dB (p7 @sp1, spnam7= rect32, spoffs7=-18257) 300u @12.9dB (p8@sp0;spnam0= sinc1.0; spoffs0 = +20219) |
| Steve Hardies, L1 p190
50 mM NaCl, 30 mM NaPi pH 5.79, T=300 |
CBCACONH (7/31)
NIH version 8.85u @-5 dB ( P1@PL1) 1035u@ 35.5 dB (P2@PL2) 47.93u@9.5dB(P30@PL30) |
||
Avance500
| Operator/sample specifics | 1H | 15N | 13C |
| Maria
Sample in 50 mM NaPi pH 6.0 |
HSQC (7/16/02)
8.45u @ -2dB (P1@PL1) 1020u @ 39dB (P2@PL2) |
HSQC (7/16/02)
45u @-3.1dB (P7@PL7) 200u @10dB (pcpd3@PL3) |
HSQC (7/16/02)
28.7u @2.35dB (P5@PL5) |
| Steve Hardies
Sample in 50 mM NaCl, 30 mM NaPi, pH 5.79 |
HSQC (7/21/02)
9.45u @ -2dB (P1@PL1) 1025u @ 38.5dB(P2@PL2) 2040 @ 38 dB (P3@sp0, with seduce1.jc) |
||