Frequencies for CBSA DRX-600 (aka
Avance 600) at 300K
Measured10/15/01 by aph.
| Spin |
Frequency (MHz) |
Approximate Chemical Shift. |
| 1H of DSS (0.0 ppm reference) |
600.0800000 (at 300K)** |
0.00 ppm by definition |
| 1H of H2O |
600.0828242* |
4.706 ppm (at 300 K) |
| 1H in center of amide HN |
600.0849207 |
8.20 ppm |
| 15N (0.0 ppm reference) (IDRR)^ |
60.8055771 (at 300K)** |
0.00 ppm by definition |
| 15N of 15N-urea |
60.8103443 |
78.4 ppm |
| 15N in center of amide 15NH |
60.8127522 |
118.0 ppm |
| 13C (0.0 ppm reference (IDRR)^ |
150.8898340 (at 300K)** |
0.00 ppm by definition |
| 2[13]C of acetate |
150.8937722 |
26.1 ppm |
| Center of Ca + Cb 13C |
150.8967749 |
46.0 ppm |
| Ca 13C |
150.8982838 |
56.0 ppm |
| carbonyl 13C |
150.9165415 |
177.0 ppm |
*This frequency is essentially termperature independent: Other
frequences unless noted will be slightly temperature dependent and should
be correced as appropriate.
**This frequency and associated chemical shift is highly temperature
dependent.
^Calculated using the IUPAC-recommended indirect referencing ratios
relative to the DSS 1H standard. The ratios are 0.101329118 for 15N
and 0.25144953 for 13C.
For setting the carrier frequencies to be used in a determination:
-
The frequency table above doesn't drift much in normal operation.
Look for it to be updated every year or so, or after the magnet is deenergized
and reenergized.
-
Expect the
-
Carrier frequencies are set in the RF channels as if the experiment was
being done at 300K.
-
Carrier frequencies are usually specified in pulse program comments in
ppm. To find the actual frequency to set in the channel for x ppm,
look up the 0.00 ref frequency above for that isotope. Then
derive the frequency to set (y) as
x = (y-ref)/ref
The most commonly specified carrier settings are already tabulated above.
For determining center point offsets in ppm to use in processing the data.
-
These should be corrected for the actual temperature of the determination.
-
In normal operation, the frequencies do not drift much
Carrier frequencies during an experiment are generally set as if the
experiment were at 300K, regardless of the actualy temperature. I.e.,
a direction to set 15N carrier at 118 ppm always means to set
it at 60.8127522. [(60.8127522 - 60.8055771)/60.8055771 = 0.000118].