Gas flow failure.
7/2/2 SCH.
Gas flow failure is indicated in the edte window as gas MISSING.
This refers to a sample temperature control system in which refrigerated
air flows through a heater to adjust its temperature and then across the
sample tube. Loss of gas flow causes the heater to automatically
kick off, and loss of temperature control at the sample. Gas flow
failure can occur either due to loss of air pressure in the university
supply lines or due to freezing up of the refrigerator unit. In the
former case, the lock lift will also not work. One can tell if there
is air pressure in the supply lines from the gauges where the supply lines
attach to the wall. There is a ball flow gauge behind the right hand
door of the acquisition computer which will also indicate loss of gas flow
specifically in the temperature control air stream.
Check the air dryer.
Blockage of the refrigerator unit may be a secondary failure due to failure
of the air dryer. The air dryer consists of a set of blue cylinders
attached to the wall inside the utility room where the Avance500 and 600
spectrometers are located. It also supplies dry air to the Avance700.
The dryer is not currently on emergency power, so if power is off for a
significant time (~12 hours), enough humid air will circulate through the
air refrigerator unit to freeze in the lines and block gas flow.
We've also seen the dryer off because someone kicked the plug out of the
wall socket. In any case, the first thing to do to correct a gas
flow problem it to check that the green power light is on on the dryer
unit. If there is no power to the wall socket that it is plugged
into, then move the plug to an active socket in the outer NMR room until
that problem can be rectified.
Thaw and dry the feed line.
The gas flow goes from the dryer to the wall gauges, then to the acquisition
computer console and through the ball gauge, then to the refrigerator unit
(1.5 x 1.5 foot box on the floor), and then through an insulated hose to
the bottom of the magnet where it connects in a cup and socket joint secured
by a pinch clamp. Turn off the refrigerator unit and disengage the
pinch clamp. After 15 to 20 minutes air flow should resume, indicated
by 525 L/hr on the edte window and by palpable air flow at the exposed
end of the delivery hose. Let dry air blow through the system to
dry it out for 3 to 4 hours before reconnecting the hose at the bottom
of the magnet and turning the refrigerator back on. You will have
to turn the heater back on in the edte window, by clicking the on/off
button next to the heater indicator.
Unresolved issues.
1) I see 525 L/hr. set now on the Avance600. Is that the nominal
rate? The ball is 1/4 to 1/3 of the way up the gauge. There
appears to be a sensor in the middle. Is this to indicate an over
pressure situation? Note that the flow rate show in the edte
window doesn't fluctuate when the flow rate starts to fail. It apparently
either reads the set value or MISSING
2) I understand that people turn it up sometimes to get stronger temp.
control. What are the acceptable limits?
3) I once experienced that someone had left the flow rate up such that
the lift failed to lower my sample. It would be best to have a nominal
rate recorded so that you could put the machine back into a known state,
however I suppose any lower number that allowed the sample to go down and
lock in but maintained constant temperature would be good enough.
I guess that people who need a higher rate turn it up after they get the
sample locked in.