These instructions are to determine a 2D 1H
-15N shift correlation spectrum (HSQC spectrum)
for a protein sample on the Bruker AMX 500 NMR spectrometer. Some
differences relevant to using the Advance 600 NMR spectrometer are noted.
These instructions assume that you sample has been isotopically labeled
by inclusion of 15N in the growth
medium, and that it has not been 13C labeled. If the
sample has been 13C labeled, see the notes included below about
calibration of the 13C 90 degree pulse that is needed to effect
decoupling of the 13Ca and
the 13CO from the backbone amide 15N. This
calibration is required prior to beginning the steps listed below.
Version 6/3/02
New reference data sets created 6/02 for the 600.
Check sfo1, sfo3 vs. frequencies posted at the console.
A skeleton of essential steps and especially important
points are in bold.
Bulleted items are intended to be explanatory or optional.
Items in brackets are unresolved issues in this draft
copy.
The protein sample should have been concentrated to 300 ul, ideally at a concentration of 1 mM. It should have been added to an 18 mm Shigemi NMR tube with the piston positioned leaving 18 mm of fluid space with no air bubbles. If the sample is to be loaded at the NMR facility, bring some of the sample buffer in case it needs to be topped off. NB! If the plunger is too low, you are destined to spend several hours trying to shim the instrument, only to have to start over by repositioning the plunger.
Problems set in gradually as the volume drops below 18 mm. An example is given of a passable shim achieved at 17 mm. Experiments can be done at 16mm or even lower, but with increasing difficulty in shimming.
A separate transmitter and receiver within the instrument will track the deuterium signal from the 5% D2O in your sample buffer. This system will detect instrument drift throughout the acquisition and make compensating adjustments.To manually set the deuterium lock:- Make sure that the LOCK key on the BSMS is off and the SWEEP key is on.
- On the xwinnmr menu bar click <Windows> and <Lockdisp> to open the lock display window.
- Type "lock d2o" to XwinNMR.
This starts a series of automated optimizations of the deuterium detection. The system then automatically engages the lock as indicated by the LOCK light on the BSMS becoming illuminated. The lock display window shows the status of the lock signal. This varies from a flat line immediately after locking to an oscillating line during signal acquisition. Some operators leave this display on the screen to assure them of a continuing strong lock signal. It can also be reopend from the menu bar at any time.Manually setting the deuterium lock:
This operation is not expected to be needed to be needed since the introduction of the lock d2o function. However, if there appears to be a failure of the automated locking function, running through the manual setting might reveal the problem.
The first section describes the general operation of the BSMS unit, which is also used in manual shimming and for pulse calibration.
General BSMS operation. This is one of several operations using the BSMS controls. The general use of the BSMS is described first, followed by the specific instructions for setting the deuterium lock.
- The keys on the BSMS are toggle switches. Pressing a key turns a function on, and then pressing it again turns it back off. There is a light on the key to indicate whether the function is on or off. A BSMS key will be indicated in caps in this document.
- Toggling on some of the keys allows you to adjust a value. Read carefully; toggling the key back off cancels the adjustment! The value is adjusted by rotating the grey knob at the bottom of the BSMS.
- Toggling on the FINE key causes the value to change more slowly as you turn the knob.
- A display in the middle of the BSMS shows the starting value on the left and the new value on the right.
- To accept the new value, press the STDBY key, or toggle on any other value setting key. This automatically turns the light out on the key you were adjusting.
- Alternatively, if you toggle off the function you were just adjusting without accepting the adjustment, then the value will revert to the starting value.
Gradshim -- a new function
was recently introduced that will use field gradients to measure the field
at different points on the Z axis and automatically adjust the Z shims,
including higher order shims, to make it homogeneous. The Avance
machines will either do shimming only on the Z axis (1D); or on all 3 axes
(3D shim). The following instructions are for 1D autoshim.
3D autoshim instructions will be added later. With the 1D autoshim
procedure, it is recommended to still read a shim file, then do the 1D
gradshim. For final shimming you will need to do the X and Y part
of the manual shimming procedure, then repeat the gradshim program.
Note: do not confuse running the gradshim program with pressing the button
on the BSMS named AUTOSHIM. We do not use the latter function.
The elementary shim operation is to depress a 2 keys
sequence on the BSMS that activates adjustment of a particular shim circuit.
The two keys will be one of the set (ON AXIS, X, Y), and one of the set
(Z0, Z1, Z2, Z3).
If properly activated, the lights on the chosen pair will be on, and the others off. The shim name will then be displayed on the BSMS display. Once the selected shim circuit is activated, turn the knob until the signal in the lock window is maximized. Changing the shim too rapidly could conceivably cause loss of the deuterium lock, in which case you should go back to the lock power and gain adjustment and reestablish the lock with AUTOLOCK. If the signal goes off the top of the screen, just decrease the lock gain or lock power. (Leave the instrument locked on during these adjustments; ie. don't touch the LOCK key).
The problem with shimming is that the different coils interact, such that optimizing one knocks the others off their optimum. For a more extensive treatment of shimming see here. For the lower order shims, one interatively optimizes them in the sequence specified below and expects to achieve a global optimum. For higher order shims (eg. Z4 and above), the system can get trapped in local optimums. For these one has to arbitrarily change the higher order shim, and then judge it by reoptimizing the lower order shim. The only interaction that might benefit by this procedure among the shims adjusted below is the Z1 Z3 pair. The higher order shims should not need readjustment if:
The 1D gradshim program automatically
handles the interactions among the z shims, including higher order z shims.
You need to be aware that there are interactions between the X and Y shims
and the Z shims. Therefore after adjusting the X and Y shims, you
should repeat gradshim.
If shimming is going badly, take the sample out and look for bubbles and such now. If you have to remove the sample from the instrument at the final shimming step, then you will have to repeat essentially the entire procedure.
Running gradshim:
Tuning requires setting a few parameters. A dataset with the appropriate parameter files has been created to set these for you. If you have not already done so, you should first create a copy of the tuning dataset in your own directory under your own ownership.
To obtain the tuning dataset from another directory:
Or -- to access the parameter file in the tune1h.xxx
dataset from your own directory:
To obtain the hsqc_fb.xxx dataset/pulse program for
the first time:
To remind yourself about what data is in a dataset, you can give [each] experiment a title.
- Type "setti" to xwinnmr. This opens the vi editor to create a file describing the dataset however you like.
- Exit the title editor with :wq
- The command "title" <yes> to xwinnmr will now display the first line of this title file below the name of the dataset in the window bar.
- ["title" <no>, however, didn't seem to remove it.]
- We are considering adding documentation to the ref datasets by this mechanism. The documentation wouldn't be copied with the dataset, but we are exploring if the command "setti" would allow you to read the documentation before you copied the dataset.
Examine the hsqc_fb pulse program comments:
Note: The Avance instrument uses sp0 instead of tp0, and spnam0 rather than tpname0.
And the commands that issue the pulse in hsqc_fb are:
The phases of the pulses that arrive at the sample must be coordinated. Their relative phases are disrupted by the different levels of power amplification. The following operation empirically determines a correction factor for each of the two soft pulses relative to the hard pulse. These correction values will be required to directly edit into hsqc_fb before it is run.
To determine the reference
phase.
To determine the relative phase of hsqc_fb pulse
p2.
To detemine relative phase of the shaped pulse
(only needed if doing flip back):
Checking the linewidth (by examining the transformed
water peak).
Contributed by Steve Hardies, based on a protocol first developed by Pete Walker, and comments by Andy Hinck.
- You will not normally have the acquisition window active at this point, but if you do you must click <return> to return to the xwinnmr main display.
- The acquisition window is active if you see the protein fid dynamically changing every few seconds on the screen.
- Leaving this display on during a long acquisition crashes the program.
- Do not close the xwinnmr window itself.
- Open a new unix shell. Go to the buttons on the desktop outside of the xwinnmr window and click <desktop><open unix shell>
- Type "xlock"; this will put up a screen saver that can only be exited with your password. This will protect your experiment from disruption by other users.
- Come back when the experiment is over.
- Use your password to exit the xlock
- Confirm that xwinnmr indicates the experiment is over.
- Use the LIFT button to retrieve your sample.
- Toggle the LIFT button off.
- Exit xwinnmr, and any other windows you have left around.
- In the unix <desktop> menu, <logout>
- Processing of the data will be described in another document. The computers in the NIS system can directly access the /u, /hinck, and /nall drives as "read only" devices by the names amx500_u, amx500_hinck, and amx500_hinck.
6/2/02 - some corrections contributed by Pete Walker were incorporated.
6/3/02 - SCH starting trying to flag some difference with the Avance
machines, but the job is not done.