;Nitrogen settings - Set carrier to amide center (ca. 118 ppm)
;p7 - square 90 15N @ pl7, for excitation,
typically p7 = 45u @ pl1 = -3.1dB
;pcpd3 - square 90 15N @ pl3, for cpd, typically pcpd3
= 200u @ pl3 = +10.0
;The following is only necessary if CARBON_LABEL is defined
;13C pulses - carrier is set according to fqlist1 frequency list
;
chose/edit in ased, should have two values,
;
first is CA (56 ppm), second is CO (176 ppm)
;p5 - square 90 13C @ pl5, for excitation,
p5 must equal 28.7u, typically pl5 = +2.6
His actual pulses:
p1 8.9u @ PL1=-2 dB
p2 = 1015u @ PL2=39 dB
p3 = 2025u @ sp0=38.5 dB
p5 = 28.7u @ PL5=2.6 dB
p7 = 45u @ PL7 = -3.1 dB ; the PL1 indication in the comments must
be wrong.
pcd3 = 200u @ PL3 = 10 dB
SF01 = 500.1323524 These all appear to be slightly wrong.
rework from the posted DSS H value.
SF02 = 125.7644958
SF03 = 50.6837118
Conflict: pulse program says use frqlist1 (sic fq1list) = 56, 176
(these are ppm, fq1list
has actual frequencies).
By my calculation below, the two values in fq1list should be:
125.7644958 (56 ppm)
125.77958311 (176 ppm)
Assuming 125.7644958 is really 56 ppm then
(125.7644958 - x)/x = 56 ppm
x = 125.7644958/1.000056 = 125.7574498155633 = carbon ref.
dss H ref = 125.7574498155633 / 0.251449530 = 500.1299855902029
SF01 in ppm would then be (500.1323524 - 500.12998559)/500.12998559
= 4.7323, which may be a temp that Andy actually used. It would correspond
to 29 C.
The second freq in the list (176 ppm) would be:
(x-125.7574498)/125.7574498 = 0.000176 = (0.000176 * 125.7574498)+125.7574498
=125.77958311
N standard should be 500.12998559 * 0.10132918 = 50.67776133
(50.67776133 * 0.000118) + 50.67776133 = 50.6837413, which is close,
but not what's set.
Working backwards: 50.6837118/1.000118 = 50.67773096
50.67773095/0.10132918 = 500.1296857 = H DSS ref.
(500.1323524 - 500.1296857)/500.1296857 = 5.332017ppm, which
seems unlikely to be a value that Andy set on purpose.
But dataset says uses ca_co.ref = O 125.76 4495.8 19712.5, there's
also a file left in the
dataset directory named fq0list with these values?
Then again F2LIST referred in the dss calibration tutorial just
has <2222222222> and item 2
in FQ1LIST is 125.76; SFO2 as set is somehow a combination of numbers
2 and 3 in ca_co.ref.
Claims ased somehow clarifies this.
OK, the command to set up such a list is edlist
O 125.76 means that the next two values are frequencies calculated
as offsets from 125.76
125.7644958 and 125.7797125 are the two frequencies set in the
list.
I wonder if the problem isn't a typo, and it's supposed to be 52
and 176 ppm?
/u/prog/docu/english/xwinproc/pdf/
F2LIST is for amx only
For the old hsqc datasets on taken on the amx, F2LIST is set to
ca_co.ref with these same contents???
FQ1LIST is for avance machines only.
Command is fqx:fy to use current position in FQxLIST and set to
channel fy.
NS was 8.
TD was 2048
As for the "optional" phcor settings, I'm going to ignore them.
I have no criteria to claim that relative phases I'd measure 1 degree differently
from the automatic phase correction
are more accurate than what the machine determined.