Chemical Shift Referencing and Temperature Calibration.

Version 6/14/02 - by Steve Hardies, from notes taken from Andy Hinck.

This document covers temperature and chemical shift calibration experiments that should be done in conjunction with protein spectra.  The document deals with correcting a systematic error that arises from the use of an external standard and the deuterium lock operating on deuterated water.  The error arises from the fact that the resonance frequency of water is temperature dependent.  The function of the deuterium lock is to compensate for instrumental drift in the field strength during and between experiments.  The lock continually alters the field to enforce a fixed resonance frequency of the internal deuterated water.  When comparing experiments done at different temperatures, the lock will have adjusted the field to cancel the intrinsic temperature shift of the deuterated water.  As a consequence, the resonance frequency for water will be falsely reported as the same at different temperatures, and the frequencies of all other signals will have an equal and opposite shift to that of water superimposed on their intrinsic temperature shifts.

To contrast with the problem created by external referencing, the temperature shift artifact does not affect internally referenced experiments.  For internal referencing, the standard 2,2-dimethylsilapentane-5-sulfonic acid (DSS) is included within the protein sample.  DSS has no intrinsic temperature (or pH) shift.  When comparing experiments at different temperatures, the DSS peak will artifactually be shifted along with the other signals by the operation of the deuterium lock.  However, during processing, the DSS peak of each experiment will be manually set as the zero point.  This will automatically subtract the artifactual shift on the 1H axis from all other signals.  There may be shifts remaining for the other proton signals.  But these residual shifts reflect true and intrinsic properties of those nuclei, and there is no intention of removing those shifts from the data.  For 13C and 15N, there is generally no corresponding internal reference molecule.  An artificial internal zero point for the 13C and 15N axes is created by multiplying the DSS frequency by a characteristic ratio (see below).  Hence these reference points will indirectly be adjusted to subtract the deuterium lock artifact from the 13C and 15N axes.

However, most experiments are not internally referenced.  This is due to the fear that the DSS may bind to the protein and alter some of its signals.  Instead most experiments are externally referenced.  During external referencing there is no peak to manually set to the zero point.  Instead, the carrier frequency, which is at the center point of the spectrum, is initially taken as the zero point.  Then an offset is added that would cause the DSS peak to become zero if it were in the spectrum.  That requires knowing what the difference between the carrier and the DSS frequency should be under the conditions of each experiment.  The carrier frequency is set to the constant water resonance of water that is enforced by the deuterium lock  The deuterium lock will cause the apparent frequency where DSS would resonate if it were there to shift with temperature.  Hence one needs to know the apparent DSS frequency as a function of temperature in order to correctly process the data.

Temperature shifts in Hz are generally divided by the frequency of the zero point reference for that nucleus and expressed in ppm.  Temperature shifts expressed in ppm are the same when measured on instruments with different field strengths, and are also in the same units as the chemical shifts they are being used to correct.  See examples at the bottom.

There are two experimental issues involved in this problem.

  1. An experiment is described that measures how close the thermistor that reports sample temperature is to the true temperature.  It consists of taking the spectrum of ethylene glycol, in which there are two peaks whose separation is highly temperature sensitive.  The thermistor can be off by several degrees C.  Obviously one can't choose the right temperature correction without knowing the true temperature of the experiment.  Best practice is to measure the offset of the thermistor from the true temperature first, and then set the experimental temperature to compensate.  This calibration is recommended to be performed about once every two weeks.  As a note on record keeping, I recommend using the settitle command during experiment set up to record the true temperature.  Otherwise the temperature correction factor tends to get misplaced before the data gets processed.
  2. An experiment is described that measures the shift between DSS and water at a particular temperature.  These shifts should come out to be 4.742 ppm at 300K, with a temperature coefficient of - 0.0119 ppm/degree (Wishart et al., J. Biomol. NMR 6: 135-140).  There is also a 0.002 ppm shift per unit pH, and possibly salt effects.  To get the most accurate possible correction, one should put DSS into your experimental buffer and make the determination.  However, assuming an experimental precision on the order of 0.01 ppm for 1H, only the temperature correction is relevant under typical experimental conditions.  Hence we use a standard consisting of DSS in water.  Doing this calibration instead of using the theoretical values is mainly an exercise in determining your own experimental accuracy.

Temperature calibration with ethylene glycol.

A sample of 100% ethylene glycol in a  wilmad tube (round bottom NMR tube) is found in that flask full of standards at the operator's table.

DSS standard.

Note that DSS is the primary standard used in protein work, but a different but related standard (tetramethylsilane (TMS)) is the IUPAC standard for proton NMR.  To distinguish the difference, chemical shifts calibrated against DSS are denoted as dDSS or d(DSS) (that's a lower case delta).  See John L. Markley et al., 1998. "Recommendations for the presentation of NMR structures of proteins and nucleic acids". Pure  Appl. Chem., Vol 70 117-142.  The Markley et al. document is posted at http://www.bmrb.wisc.edu/home/iupac.pdf.

A DSS sample in deuterated water is found in a wilmad tube in the flask of standards on the operator's table.


This page was last updated June 14, 2002


[Home]    [Spectrometers]    [User Guidelines]    [Scheduling]    [UNIX Software Information]    [Fees]
   [Education and USER BASE]    [NMR Links]    [Contact]     [Center for Biomolecular Structure Analysis]
   [UTHSCSA]     [Department of Biochemistry]     [Disclaimer]     [Copyright]