Tuning (15):
SCH - 7/24/02
The 1H RF coil needs tuned each time a sample is placed in
the spectrometer. Tuning of the 1H RF coil should be repeated
if the same sample is removed and then returned to the spectrometer.
The 13C and 15N coils are less sensitive to the specifics
of the sample, but still require occasional tuning. There is some
risk that a previous user tuned the 13C and 15N
coils and did it badly. Tuning for the avance 500 and 700 is done
entirely from the console using the atmm command. Tuning of
the Avance600 is done by adjusting pots on the bottom of the magnet after
using the wobb command (1). Tuning of the 15N
coil of the Avance600 only requires changing a cable. You should
consider rechecking 1H tuning just prior to starting an experiment
with a lengthy set up (13).
Tuning with atmm command.
-
If you do not have a copy under your own ownership of the relevant tuning
dataset (tune1h, tune13c, tune15n), obtain it from the ref user using edc
(6).
-
Type atmm (5). When the message line indicates that
the display is ready in the acquisition window, type acqu.
-
Click the <wobb-SW> button (mid left), agree with the two popup boxes,
and wait for the curve to redisplay (2,7,8).
-
Use the tuning adjustments in the popup window to center the tip of the
V-shaped portion of the Wobb curve (4,11) on the solid vertical line, and
then use the matching adjustments to lower the tip of the V as far as it
will go (10).
-
You can use the /8 or /2 buttons (upper left) or equivalent commands to
see the tip of the V better (9).
-
Press and hold the left mouse button down on the adjustment buttons, particularly
the fine adjustment (indicated by single arrows).
-
The tuning and matching adjustments interact, so you may have to go back
and forth between them (12).
-
Exit the atmm routine by <file> <exit> in the tuning adjustment
window; then click <return> in the xwinnmr window (3).
Tuning with wobb and manual adjustment of the capacitors.
-
If you do not have a copy under your own ownership of the relevant tuning
dataset (tun1h, tune13c, tune15n), obtain it from the ref user using edc
(6).
-
Type wobb (5). Wait for the message line to announce that
the wobb curve is in the acquisition window, then type acqu.
-
If the wobb curve already centered on the solid vertical line and touching
the bottom, tuning is not needed; type halt.
-
To tune the probe, you will have to go to the magnet. Since you can not
see the console from there, there is an LCD display to guide the operation
on the preamp, which is located on the floor behind the magnet.
-
Locate two capacitors for the coil to be tuned on the probe located at
the underside of the magnet. The color code for each nucleus is indicated
where the cables attach to the probehead unit. The tuning and matching
capacitors are distinguished by a T or M on the capacitor shank.
-
Using the tool connected by a chain, turn the pot marked T until the lit
up LCDs in the horizontal row are centered. The capacitor should turn
easily. Do not force it. If it doesn't turn, you have reached
the endpoint (10). Never force the capacitor past this point, else
you will damage it and the probe will have to be sent back to Bruker for
repair.
-
Then turn the pot marked M to reduce the height of the vertical row of
lit LCDs.
-
The two pots interact, so go back and forth between them optimizing the
tuning (12).
When optimally tuned, there should be a single LCD lit in the center
of the horizontal row and two or three lit at the bottom of the vertical
row.
Return to the terminal and type "halt" to stop the acquisition ("halt"
stops the acquisition after that pulse is finished.)
-
Click <return> on the acqu menu to exit to the main xwinnmr window (3).
Tuning the 15N coil on the Avance600
Tuning of the 15N probe requires the following additional steps
on the Avance 600 (4):
-
Locate where a cable from the probehead passes through a small box labeled
"13C bandpass" which screws into the second unit from the top
in the preamplifier stack. Unscrew the 13C bandpass filter
from the preamp unit.
-
Locate where a cable from the probehead passes through a small box labeled
"15N bandpass" and connects to another cable that bypasses the preamp unit.
Unscrew the side of the 15N bandpass filter away from the probe
and connect it to the preamp unit where the 13C connection had
been. Be very gentle with the 15N and 13C
bandpass filters; they contain inductors (essentially small coils of copper
wire) that are sensitive to mechanical distortion which in turn will alter
their filtering properties.
-
Complete the 15N tuning operation in this configuration.
-
Return the cables to their original configuration before attempting any
other operation (4a).
Comments and Troubleshooting.
-
(1) More exactly, it is the probehead installed that determines if atmm
or wobb is used. See the instrument status page for information
on the probeheads available for each spectrometer. If servo motors
for tuning by the atmm process are installed on the probehead, then
there are no adjustment pots and manual tuning at the underside of the
magnet is not possible. The gradshim window gives the name
of the installed probehead.
-
(2) The <wobb-SW> button (or wbchan command) reinitializes the
sweep width from a parameter in the dataset. The wobb function
(in manual tuning) can tolerate different sweep widths, but the atmm
function requires a particular sweep width for each nucleus. At present,
atmm is sometimes starting with an inappropriate sweep width, or
at least an inaccurate display suggesting an inaccurate sweep width.
A future fix may make this step unnecessary. The displayed frequency
range should equal the sweep width, which should be 40 MHz for 1H
and 1 MHz for 13C or 15N. We have observed
the 1H display to be far enough obscured by the errant sweep
width setting to place the resonance position off screen.
-
(3) If atmm or wobb is not properly exited the process
continues in the background without display window. Then you
will get a "device in use by other process" error when you subsequently
try to do another acquisition. If so, use the kill command
to kill the persisting atmm process.
-
(4) The cable change for tuning 15N on the Avance500 is required
because its preamp can only handle only one X channel at a time (i.e. 15N
or 13C). The system is usually set up to route the 13C
channel through the pre amplifier. Thus, for 15N tuning,
the 15N signal has to be rerouted through one of the pre amplifier channel
usually used for 13C. This is partly accomplished by the
cable switch. Additionally, there are parameters in the tune15n data
set which reroute the RF so that 15N RF enters the rear of the
preamplifier instead of 13C RF.. To see how signals are
routed by dataset parameters use the edasp command and compare the
tune15n data set to other data sets. The Avance 500 and 700 spectrometers
are equipped with two X-nucleus preamplifier units each, thus no cable
change is necessary.
-
(4a) Do the 15N tuning before the 13C tuning.
That way, if you forget to change the cables back, you will immediately
notice it.
-
(5) A message to the effect that a nucleus can not be wobbled means that
you probably haven't loaded a tuning dataset.
-
(6) An error message to the effect that write permission is denied may
result if you have not established ownership of the tuning dataset. There
was at one time a bug in the setup of the avance 500 such that edc immediately
gave a "permission denied" error when acquiring a ref dataset, even though
there was in fact no error. This problem should now be fixed.
There is a "ref" user on both drives /u and /avance6001 on the Avance600,
but only the one on /avance6001 has the reference library.
-
(7) If in doubt about the shape of the curve, click <wobb-SW> and agree
to the two popup boxes.. The keyboard equivalent is wbchan.
It does not help to type wbchan prior to running atmm.
-
(8) If you may have inadvertently changed parameters in the tuning dataset
by typing something in the popup boxes generated by <wobb-SW>, then
the best thing to do is to reobtain the tuning dataset from the ref library.
-
(9) If vertical rescaling has confused the issue, restore default vertical
scaling by clicking the up arrow pointing to the horizontal line in the
upper left button panel. The keyboard equivalent is vreset.
-
(10) If tuning and matching are both far off to start with, and tuning
is not working well, then try improving the matching first.
-
(11) The 40 MHz sweep width used by atmm for the 1H probe
includes secondary minima at the high and low end of the displayed frequency
range. This is normal. Different sweep widths are set up for
the atmm function than for the wobb function where atmm is
not implemented. Therefore it is normal for the curves to appear
to be different.
-
(12) The 1H wobb curve on both the avance500 and avance600 has
a distortion that places two closely spaced minima right at the resonance
frequency when the coils are tuned. The distortion also makes the
tip of the curve jump around a little bit making it hard to decide on optimal
settings. This does not indicate an error in your technique.
If you are that close, your are close enough.
-
(13) If 1H tuning drifts over the course of the experiment,
it may because of an inhomogeneity in the salt concentration in the sample
tube. For example, if you top off a sample with water and don't mix
it up, this behavior may result. Bubble formation is another possibility.
-
(14) Resonance frequencies currently in the Avance500 tuning datasets are
500.13385041, 125.757789, and 50.677733. For the Avance600, they
are 600.084620616, 150.8946364, and 60.820685787851. These don't
have to be fine adjusted to the posted reference frequencies.
Theory
-
(15) Tuning actually refers to two processes, called tuning and matching.
Tuning is adjusting the resonance frequency of the RF coil to more closely
match the frequency range of the nucleus under observation. The tuning
dataset should have the relevant frequency set as an sfon
parameter within it (14). Matching refers to adjusting the impedance
of the RF coil to that of a reference resistor embedded in the electronics
to represent an optimal operating condition. The resonance frequency
arises as follows: As the transmitter passes and alternating current
through the coil, the associated electromagnetic fields interact most constructively
at a particular frequency based on the geometry of the coil. Impedance
arises because the generated coil fields resist passage of alternating
current. Impedance is lowest at the resonance frequency, and the
intensity of the emitted RF field is highest in the center of the coil
at the resonance frequency. Working in conjunction with the receiver,
the coil is most sensitive at generating an alternating current from radiation
emitted by the sample if the radiation is near the resonance frequency.
The impedance of the RF coil includes a component of inductance contributed
by salt ions in the sample itself. The resonance
frequency of the coil can be somewhat modified by a variable capacitor
placed between the leads of the coil. This is the tuning capacitor.
The impedance can be modified by a second capacitor within one of the leads
of the coil. This is the matching capacitor. During the tuning
operation, the transmitter synthesizes a sweep through the frequency range
by repeatedly putting out a series of closely spaced pulses. The
wobb window display acts as an oscilloscope in in the circuit in sync with
the transmitter. The dip in the wobb curve represents the resonance
frequency as the point of lowest impedance. A device called a bridge
balances the current through the coil with a shunt through a reference
resistor. The lowest overall impedance occurs when the impedance
of the coil matches that of the reference resistor. This allows the
effect of adjusting the matching capacitor to be visualized as vertical
motion of the dip in the wobb curve.
-
Further discussion can be found the Bruker acquall.pdf file in /u/prog/docu/english/xwinproc/pdf/
or through the xwinnmr help function.
-
In certain circumstances, the 1H coil is intentionally detuned
(frequency adjusted to one of the highest points on the Wobb curve) to
eliminate radiation damping.